Installation of the Optional Sample Cooler or Thermostat
Install the Sample Cooler or Thermostat
5
Slide in the sample cooler or thermostat the halfway.
94
WA R N I N G
Module is partially energized when switched off, as long
as the power cord is plugged in.
Repair work at the module can lead to personal injuries,
e.g. shock hazard, when the cover is opened and the
module is connected to power.
Make sure that it is always possible to access the
power plug.
Do not use the sample cooler or thermostat if it is not
operating correctly or has been damaged. Disconnect it
from the power supply and call your local service
center.
Remove the power cable from the module before
opening the cover.
Do not connect the power cable to the module while
the covers are removed.
If the sample cooler or thermostat is disconnected from
the power supply, you should wait for at least five
minutes before switching on the compressor.
C AU T I O N
Damaged electronics
To avoid damages of the electronics of the module
make sure the power cords are unplugged before
disconnecting or reconnecting the sampler to the
sample cooler/thermostat cables.
6
Connect power cable and signal/data cable.