Install the stator ring with the short of the two pins
facing towards you at the 12 o'clock position. Ensure the
ring sits flat on the valve body.
Agilent InfinityLab LC Series Vialsamplers User Manual
8
Install the stator head. Tighten the bolts alternately two
turns at a time until the stator head is secure.
N O T E
Do not over-tighten the screws. The screws hold the
assembly together and do not affect the sealing force.
The sealing force is automatically set as the screws
close the stator head against the valve body.
Maintenance and Repair
Exchange the Rotor Seal