Installation of the Optional Sample Cooler or Thermostat
Install the Sample Cooler or Thermostat
Condensate inside the sample cooler or thermostat
C AU T I O N
Damage to the electronics of the module
In the event of a damage
WA R N I N G
Do not open the Sample Thermostat. There are no serviceable parts inside.
N O T E
If the sample cooler or thermostat is disconnected from the power supply, you should wait
N O T E
for at least five minutes before replugging and switching on the compressor again.
Even under average humidity conditions, a significant amount of condensed water gathers
N O T E
every day. A suitable container must be provided and emptied regularly in order to avoid
overflow.
For best performance of the sample cooler or thermostat all drawers must be installed in
N O T E
the sampler.
For the sample cooler installation in a sampler the serial number of the sample cooler
N O T E
must be DEBAT02001 or higher.
Depending on the ambient conditions in the lab, the amount of condensate can vary from
N O T E
200 mL to 2 L per day. Waste containers for the condensate should not be filled to the top.
Regularly empty the waste container.
92
After installation of the sample cooler or thermostat, wait at least 30 min before
switching on the module.
Make sure there is no condensate inside the module.
Keep open fire or sources of ignition away from the device.
Ventilate the room for several minutes.
Do not use the Sample Thermostat any more.
Agilent InfinityLab LC Series Vialsamplers User Manual