Preparing the Module
Installing Capillaries
3
Carefully slide the ferrule components on after the nut
and then finger-tighten the assembly while ensuring that
the tubing is completely seated in the bottom of the end
fitting.
5
Loosen the nut and verify that the ferrule is correctly
positioned on the tubing or capillary.
The first time that the swagelock fitting is used on a column or an injection valve, the
N O T E
position of the ferrule is permanently set. If changing from a column or an injection valve
to another, the fitting may leak or decrease the quality of the separation by contributing to
band broadening.
116
4
Use a column or injection valve to gently tighten the
fitting which forces the ferrule to seat onto the tubing or
capillary.
N O T E
Don't overtighten. Overtightening will shorten the
lifetime of the fitting.
Agilent InfinityLab LC Series Vialsamplers User Manual