12
Hardware Information
Firmware Description
Firmware Updates
Firmware updates can be done with the Agilent Lab Advisor software with
files on the hard disk (latest version should be used).
Required tools, firmware and documentation are available from the Agilent
web:
http://www.agilent.com/en-us/firmwareDownload?whid=69761
The file naming conventions are:
PPPP_RVVV_XXX.dlb, where
• PPPP is the product number, for example, 1315B for the G1315B DAD,
• R the firmware revision, for example, A for G1315B or B for the G1315C
DAD,
• VVV is the revision number, for example 650 is revision 6.50,
• XXX is the build number of the firmware.
For instructions on firmware updates refer to section Replacing Firmware in
chapter "Maintenance" or use the documentation provided with the
Firmware Update Tools.
Update of main system can be done in the resident system only. Update of the resident
N O T E
system can be done in the main system only.
Main and resident firmware must be from the same set.
Figure 39
Firmware Update Mechanism
Agilent InfinityLab LC Series Vialsamplers User Manual
265