Agilent Technologies InfinityLab LC Series User Manual page 123

Vialsamplers
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

Table 13
Control settings
Figure 22
Control settings
For additional help and support, highlight the desired area and press the F1 key. A help
N O T E
screen will open with additional information and documentation about the topic.
Agilent InfinityLab LC Series Vialsamplers User Manual
The Sampler control parameters are in the following sections:
• Missing Vial
Mark the Ignore missing vial check box to specify that, if a vial is
missing, the injector ignores it and continues with a 6-second
dummy run. The message "Missing vial <x>" is logged, and the
system continues with the next injection.
• Illumination
Toggles the illumination of the sample area, On or Off.
• At Power On
The section is available when a cooler or thermostat is installed and
configured. Mark the Turn on Thermostat check box to specify that
the cooler/thermostat is switched on automatically when the
instrument is switched on.
• Thermostat
The section is available when a cooler or thermostat is installed and
configured and the Constant temperature mode is selected.
Select On to switch on the cooler/thermostat. Specify the required
temperature in the adjacent field. The specified temperature must
be at least 5 °C below ambient for proper temperature control.
Select Off to switch off the cooler/thermostat.
• Automatic Turn On
You can set a date and time at which the cooler/thermostat
switches on automatically.
• Pump connected to Sampler
Use this section to specify the pump that is used with the
Vialsampler. If more than one pump is configured, display the
drop-down list and select the appropriate pump from the list.
• Enable Analysis
This section is available when you have a thermostated autosampler
with a thermostat element connected and enabled only when the
Thermostat controlled by method check box in the Options section
of the Configuration dialog box is cleared. It indicates the
temperature conditions under which an analysis will be made. Select
the With Any Temperature option to specify that the analysis can be
started irrespective of the temperature of the sample. Select the
Temperature within +/- 1°C option to specify that the analysis can
be started only when the temperature of the sample is within 1 °C of
the specified value.
N O T E
The Enable Analysis section is disabled when Not controlled is
selected in the Temperature section.
Preparing the Module
Setting up the Vialsampler
5
123