Agilent Technologies InfinityLab LC Series User Manual page 203

Vialsamplers
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

1
In the Local Controller start the maintenance mode and
select Change loop function.
OR
In the Agilent Lab Advisor software select Service &
Diagnostics in the system screen Maintenance
Positions > Change loop, click Start and wait until the
needle assembly is in maintenance position.
2
Open the front door.
WA R N I N G
Sharp needle
Uncovered needles may cause injuries
Do not touch the tip of the needle.
4
Remove the safety cover.
Agilent InfinityLab LC Series Vialsamplers User Manual
Exchange the Sample Loop Assembly
3
Unscrew the safety cover screw.
5
Check if the needle is in maintenance position.
Maintenance and Repair
9
203