Removing the 500 Circuit Board Assembly

For this procedure the processor must be deenergized. See page 1-7. In addition, the DRYER
PANEL and the RECEIVING END ACCESS PANEL must be removed from the processor. See
page 1-3 for instructions, if necessary.
[1]
Move the MAIN CIRCUIT BREAKER CB1 to the ''O'' position to deenergize the processor.
(a)
Place the ESD MATT on a flat surface. Use GROUNDING KIT TL-3346.
[2]
Locate the 500 CIRCUIT BOARD ASSEMBLY in the ELECTRICAL BOX. See Figure 6-15.
246616
IMPORTANT
Possible damage from electrostatic discharge.
6-19