Removing the Blower Assembly

IMPORTANT
For this procedure the processor must
be deenergized. See page 1-7. In
addition, the DRIVE SIDE ACCESS
PANEL must be removed from the
processor. See pages 1-3 through 1-5
for instructions, if necessary.
[1]
Remove the COVER from the BLOWER
DRIVE MOTOR.
[2]
Remove the 3 MOUNTING BOLTS from the
BLOWER ASSEMBLY and BLOWER DRIVE
BELT.
[3]
Remove the BLOWER ASSEMBLY.
[4]
Loosen the 2 SETSCREWS to remove the
BLOWER DRIVE BELT PULLEY from the
BLOWER ASSEMBLY.
[5]
Install the BLOWER DRIVE BELT PULLEY
onto the new BLOWER ASSEMBLY.
[6]
Install the new BLOWER ASSEMBLY using
the 3 MOUNTING BOLTS removed in step 2.
[7]
For correct operation of the BLOWER DRIVE
BELT, align the BLOWER DRIVE MOTOR
and the BLOWER DRIVE BELT PULLEYS.
[8]
Install the BLOWER DRIVE BELT.
Correct tension is achieved when the
BLOWER DRIVE BELT does not make
loud noises when you energize the
processor. This step requires that you
energize the processor in order to test
for correct tension, and then
deenergize the processor to correct the
tension.
[9]
Adjust the position of the BLOWER DRIVE
MOTOR to obtain the correct tension of the
BLOWER DRIVE BELT.
[10]
Install the BLOWER MOTOR COVER.
4-2
INCORRECT
BLOWER
ASSEMBLY
MOUNTING
BOLT (3)
Figure 4-1 Removing the Blower Assembly
CORRECT
DRIVE PULLEY
SETSCREWS
BLOWER
DRIVE
BELT
BLOWER
DRIVE
MOTOR
MOTOR
MOUNTING
BOLT (3)
H108_0074CCB
H108_0074CA
246616