Adjusting the Entrance Detector Switch

The following procedure does not apply if you have a MULTILOADER or an AUTOMATIC FILM FEEDER
connected to the processor. Call Service Support if you need another version of this adjustment procedure.
If a portable computer is used, see the procedure ''Using a Portable Computer to Adjust the
Entrance Detector Switch'' on page 6-8.
[1]
Move the MAIN CIRCUIT BREAKER CB1 to the ''I'' position to energize the processor.
[2]
Remove the TOP COVER and DRIVE SIDE ACCESS PANEL. If necessary, see page 1-3.
[3]
Set the processor to the Standard Cycle. See the Operator Manual.
[4]
Make a FILM TOOL. Use the following specifications:
2.54 cm (1.0 in.)
5.08 cm (2.0 in.)
5.08 cm (2.0 in.)
30.48 cm (12 in.)
246616
IMPORTANT
Remove this strip and discard.
Remove this strip and save.
Film area that must be under the LED's.
Green (Unexposed)
Radiographic Film
43.18 cm (17 in.)
Figure 6-2 Making a Film Tool
10.16 cm (4.0 in.)
25.40 cm (10 in.)
H108_0168DC
6-3