Removing the Recirculation Pump

IMPORTANT
For this procedure the processor must
be deenergized. See page 1-7. In
addition, the NON DRIVE SIDE
ACCESS PANEL must be removed
from the processor. See pages 1-3
through 1-5 for instructions, if
necessary.
[1]
Disconnect CONNECTOR P/J 7.
[2]
Remove the SPLASH GUARD from over the
RECIRCULATION PUMP.
[3]
Place a CLAMP, TL-2170, on each of the 4
RECIRCULATION TUBES to prevent leakage
of replenishment solution.
[4]
Disconnect and label the 4 RECIRCULATION
TUBES:
• DEV IN
• DEV OUT
• FIX IN
• FIX OUT
[5]
Remove:
• 2 MOUNTING BOLTS
• RECIRCULATION PUMP.
[6]
Install the new RECIRCULATION PUMP.
[7]
Install all parts removed in previous steps.
[8]
Connect:
• 4 RECIRCULATION TUBES
• CONNECTOR P/J 7
[9]
Remove the CLAMPS TL-2170.
[10]
Energize the processor.
[11]
To check for recirculation, observe the surface
of the solution for movement.
5-2
FIX IN
FIX OUT
RECIRCULATION
PUMP
MOUNTING
BOLT (2)
CONNECTOR J7
Figure 5-1 Removing the Recirculation Pump
SPLASH
GUARD
TUBING
DEV IN
DEV OUT
H108_0078CCB