Periodic Maintenance; Cleaning Rollers, Racks, Crossovers, And Tanks - Kodak X-Omat 480 RA Service Manual

Hide thumbs Also See for X-Omat 480 RA:
Table of Contents

Advertisement

Periodic Maintenance

For these procedures the processor must be deenergized. See page 1-7 if necessary.
[1]
Once a month, or after processing 5,000 sheets of film, calibrate the REPLENISHER PUMPS. See
section ''Calibration of the Replenishment System'' in the Operator Manual.
[2]
Check that the MAIN DRIVE CHAIN is lubricated.
• If the DRIVE CHAIN is dry, apply lubricant to the DRIVE CHAIN. Use NLG1-No. 2 Lithium Ball and
Roller Bearing Grease TL-2324.
• If the DRIVE CHAIN is rusty, remove it and install a new one.
• Wipe the ENTRANCE SLOT of the 5600 CIRCUIT BOARD ASSEMBLY with a damp cloth.

Cleaning Rollers, Racks, Crossovers, and Tanks

Mix the Kodak System Cleaner as directed. Wear rubber gloves, safety glasses, and protective
clothing.
Do not place the RACKS or CROSSOVER ASSEMBLIES in Kodak System Cleaner.
[1]
Once every 3 months, or as experience indicates, clean the DEVELOPER RACK ROLLERS and
DEVELOPER TANK using Kodak Developer System Cleaner and a synthetic sponge. To access the
inner ROLLERS on the RACK, you must remove all of the outer ROLLERS except for the bottom one.
To remove the outer ROLLERS, you must first remove the STUDS from the drive side of the RACK.
[2]
Clean the FIXER RACK and FIXER TANK with Kodak Fixer/Wash System Cleaner.
[3]
Clean the WASH RACK with Kodak Fixer/Wash System Cleaner or a mild solution of chlorine bleach if
biological growth exists. Use 2 fluid ounces (60 mL) of bleach for every 1 gallon (3.8 L) of water.
To prevent chemical contamination, rinse ROLLERS thoroughly with warm water to remove all the
System Cleaner.
[4]
Rinse all ROLLERS and RACKS thoroughly with warm water to remove all of the System Cleaner.
[5]
Remove, clean, and install the LEVEL SENSING PROBES.
7-4
IMPORTANT
246616