Removing the Wash Drain Solenoid

For this procedure the processor must be deenergized. See page 1-7. In addition, the TOP
COVER and the NON-DRIVE SIDE ACCESS PANEL must be removed from the processor. See
pages 1-3 through 1-7 for instructions, if necessary. Also make sure that the water supply to the
processor is shut off for this procedure.
[1]
Disconnect the 2 TERMINALS. See Figure 5-25.
[2]
Loosen the 2 HOSE CLAMPS.
[3]
Remove the 2 HOSES.
As you remove these HOSES, place a mark on each HOSE so that you can identify the HOSES for assembly.
[4]
Remove the 2 SCREWS.
[5]
Remove the WASH SOLENOID from the processor.
[6]
Install the new WASH SOLENOID.
[7]
Install all parts removed in previous steps and make the necessary connections.