Removing the Jet Weaver

Tools required
p/n
8710-0510
8710-0899
Parts required
#
2
Preparations
• Select Do not use mixer in ChemStation.
• Switch off the pump at the main power switch.
1
Remove capillary connections from the Jet Weaver to
the Multi Purpose Valve.
1290 Infinity Quaternary Pump User Manual
Description
Open-end wrench 1/4 — 5/16 inch
Screwdriver Pozidrive Shaft
p/n
Description
0100-1259
Plastic fittings

Removing the Jet Weaver

2
Open the screw, which fixes the Jet Weaver to the front
panel.
9