Table C‐3. Transducer Block Parameter Definitions
Label
PARAMETER_NAME
INST_ALERTS_CONFIG4
Cycle Count Alert Point
CYCLE_COUNT_ALRT_PT
Cycle Count Deadband
CYCLE_COUNT_DB
Travel Open Alert Point
TVL_OPEN_ALRT_PT
Travel Open Deadband
TVL_OPEN_DB
INST_ALERTS_CONFIG5
Travel Closed Alert Point
TVL_CLOSED_ALRT_PT
Travel Closed Deadband
TVL_CLOSED_DB
Travel Lo Lo Alert Point
TVL_LO_LO_ALRT_PT
Travel Lo Lo Deadband
TVL_LO_LO_DB
INST_ALERTS_CONFIG6
Travel Lo Alert Point
TVL_LO_ALRT_PT
Travel Lo Deadband
TVL_LO_DB
Travel Hi Alert Point
TVL_HI_ALRT_PT
Travel Hi Deadband
TVL_HI_DB
INST_ALERTS_CONFIG7
Travel Hi Hi Alert Point
TVL_HI_HI_ALRT_PT
Travel Hi Hi Deadband
TVL_HI_HI_DB
AUX_MIN_SHORT_ALRT_TIME
AUX_MIN_CLEAR_TIME
Index
RO /
Mode
Number
RW
130
> = 0
130.1
RW
ALL
130.2
RW
ALL
0 - 100%
130.3
RW
ALL
-25% to 125%
> = 0%, < =100%
130.4
RW
ALL
131
131.1
RW
ALL
-25% to 125%
> = 0%, < = 100%
131.2
RW
ALL
131.3
RW
ALL
-25% to 125%
> = 0%, < = 100%
131.4
RW
ALL
132
132.1
RW
ALL
-25% to 125%
> = 0%, < = 100%
132.2
RW
ALL
132.3
RW
ALL
-25% to 125%
> = 0%, < = 100%
132.4
RW
ALL
133
133.1
RW
ALL
-25% to 125%
> = 0%, < = 100%
133.2
RW
ALL
133.3
RW
ALL
>= 0, <=120 seconds
133.4
RW
ALL
>= 0, <=120 seconds
Range
Default Value
1,000,000%
1%
99.5%
1%
0.5%
5%
-25%
5%
-25%
5%
125%
5%
125%
5%
20 sec
3 sec
Blocks— Transducer Block
Protect
Description
Category
Data Type: Uint32
Alerts when number of cycles is too
SETUP
large.
Data Type: Float
SETUP
Deadband
Data Type: Float
SETUP
Alert when valve is open.
Data Type: Float
SETUP
Deadband
Data Type: Float
SETUP
Alerts when valve closed.
Data Type: Float
SETUP
Deadband
Data Type: Float
SETUP
Alert when valve position is less than
alert point.
Data Type: Float
SETUP
Deadband
Data Type: Float
Alert when valve position is less than
SETUP
alert point.
Data Type: Float
SETUP
Deadband
Data Type: Float
Alert when valve position is greater
SETUP
than alert point.
Data Type: Float
SETUP
Deadband
Data Type: Float
SETUP
Alert when valve position is greater
than alert point.
Data Type: Float
SETUP
Deadband
Data Type: Float
Time required for the auxiliary
SETUP
terminal to be shorted before alert
goes active.
Data Type: Float
SETUP
Time required before alert, currently
active, to become clear.
June 2017