19. HARDWARE CHECK
This check is only available for the remote contacts (option). It checks the input signals (shorted or open) at the
remote contacts terminals.
Check Flow Chart
Operation screen
[Engineering mode selection]
 
  
Cursor
[Selecting check]
 
  
[Displaying check 1]
 
   
Cursor
[Displaying check 2]
 
  
Cursor
[Selecting remote contacts]
 
  
Cursor
[Selecting input signal]
 
 
SHIFT
A - Z
3 sec. or more
Check
°F
°C
Cursor
/
%
(1)
(2)
(3)
°F
°C
ENTRY
EX1
EX2
EX3
EX4
ENTRY
-115-

19.6 Remote Contacts Input Check

Remote contacts terminals
DISP
2 sec. or more
6
ENTRY
Remote contacts signals and functions
Eleven kinds of operations are
Ref. 1
available by the remote contacts
signals. The operation allocated to
Ref. 2
each terminal (EX1 to EX4) can be
checked. ⇒ See Section 13.2.
Reference 1
Press
Check 1 and 2.
Reference 2
Ref. 3
(1)  : Alarm output check
(2)  : Remote contacts input check
(3)  : Communications interface check
Reference 3
The input signals at EX1 to EX4 are:
0: Open
EX1 EX2 EX3 EX4
COM
Check 1 and 2 switching
or
to switch
/
%
Types of check 2
Input signal
1: Shorted