Copying To Other Channels - Omega RD2800 Series User Manual

Hybrid recorders
Table of Contents

Advertisement

11. OTHER PROGRAMMING
For the parameters, such as [range/printing range], [scale], [engineering unit] and [tag], which are needed to be
programmed in each channel, the parameters of the specific channel as a reference channel can be copied to a
desired channel.
1
Programming Mode ...

Reference
(Fixed)
2
Programming Flow Chart
<Example> Copying of parameters for [range/printing range] in CH 1 to CH 2 to 4.
Operation screen
[Check]

Channel
[To be ready for programming]

Cursor
[To be ready for copying]
 
[ Programming reference channel ]
  
Reference channel
[ Programming copying channel ]
  
YES
Error message appears?
Keys other than
SHIFT
Example of [range/printing range]

Copy channels
channel
 :
For sequential channel
・ :
For non-sequential channel
Fill unused digits with spaces.
SHIFT
RANGE
SET
END
SPACE
SHIFT
COPY(=)
Cursor
1
CHART
Various
keys
ENTRY
NO
Temporarily stored
SET
SHIFT
END
Stored
Ref. 2
-80-
11.14

Copying to Other channels

Remarks 1
DISP
7
Press
Remarks 1
any skip channel is programmed.
Reference 1
• Programming "to":
• Decimal point:
CLOCK
• Deletion of decimal point:
SPACE
COPY(=)
Reference 2
Store the [temporarily stored]
Ref. 1
parameters into memory.
(A programming change mark is
printed.)
* Fill unused digits with spaces.
Character display
Programming
lamp*
* R: Range/Printing range
S: Scale
U: Engineering unit
T: Tag
Check if any skip channel
is programmed
or
to check if
%
/
Various keys
+
SHIFT
(-)
in the next digit
in the next digit
Storing

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rd200 series

Table of Contents