13. REMOTE CONTACTS
This explanation is only for the remote contacts (option).
0
2
Press
and
CLEAR
TAG
more to display the "Operation printing position"
programming screen. Printing Nos. are A, B, C and D. The
default values are A = 20, B = 40, C = 60 and D = 80.
1
Operation Record
The programmed recording positions are recorded by
the plotter pen. When the contact signal is shorted, the
recording position shifts to 5 mm to the right from the
programmed operation record position. The record Nos.
(A to D) are also recorded at a fixed interval.
2
Programming Mode

Record No.
3
Programming Flow Chart
<Example> Setting of the record position of record No. B to 40%
Operation screen
[Check]
  
Printing
Printing No.
position
[To be ready for programming]
  
Cursor
[Selecting printing No.]
  
Printing No.
[Programming printing position]
  
Printing position
If any error occurs, press
any key other
than
.
SHIFT
Ref. 3
Programming other channels?
YES
simultaneously for 3 seconds or
Record position
0
CLEAR
TAG
3 sec. or more
SET
END
/
%
4
0
SCALE
CLEAR

ENTRY
Temporarily stored
NO
SET
SHIFT
END
Stored
Ref. 4
-91-
13.4 Programming Operation Record Position
Operation record lines and Nos.
10
A
5mm
Remarks
For operation record, the "Terminal
DISP
Allocation for Operation", allocations of
2
record Nos. (A to D) to terminal Nos.
(EX 1 to EX 4), is necessary. See
Ref. 1
Section 13.2.
Reference 1
Press
record Nos. record positions can be
checked.
°F
Reference 2
Record position is programmable from
Ref. 2
10% to 90% of the printing range.
Reference 3
Error is displayed when the
programmed value is between 0%
and 9% or between 91% and 99%.
Reference 4
Store the [temporarily stored]
parameters into memory. (Programming
change mark is printed.)
90(%
Programming range
B
Record No.
Record line when
the contact is open
Record line when the
contact is shorted
5mm
The terminal allocation
is necessary.
Other speed No. check
or
to change
/
%
Record position range
Error display
Storing

Hide quick links: