19. HARDWARE CHECK
This check is only available for the communications interface (option). It checks the communications function by
displaying the signal received after transmitting it.
1
Connections
Except for RS-485, short the transmission and reception terminals.
RS-232C
Shorted
SD
RD
SG
Short SD and RD terminals.

19.7 Communications IF Check

RS-422A
Shorted
Shorted
SDA SDB RDA RDB SG
Short SDA and RDA terminals.
Short SDB and RDB terminals.
-116-
RS-485
SA
RA
SG
Open circuit.