Microscan MS-9 User Manual page 141

Hide thumbs Also See for MS-9:
Table of Contents

Advertisement

Laser High
When enabled, a message is transmitted whenever the laser current exceeds a factory-
calibrated reference value which cannot be altered. The message repeats once every 30
minutes until the condition is corrected.
Laser High Status
Alerts the user to impending laser failure. (Contact your service repre-
Usage:
sentative.)
Enables the Laser High message.
Definition:
<K400,warning message
Serial Cmd:
novram corrupt warning status>
Default:
Enabled
0 = Disabled
Options:
Laser High Message
Definition:
Defines the Laser High message.
<K404,laser high
Serial Cmd:
Default:
HIGH_LASER
Any 1 to 10 character ASCII string except NUL, <, or >.
Options:
status,laser high
1 = Enabled
message>