Sequence On Noread

Sequence On Noread is useful when the scanner needs to stay in
Usage:
sequence even if no decode occurs.
When Sequence On Noread is Enabled and Matchcode is set to
Definition:
Sequential, the scanner sequences on every match or noread. When
disabled, it does not sequence on a noread.
<K223,matchcode type,sequential matching,match start position,match
Serial Cmd:
length,wild card
match>
Default:
Enabled
Options:
0 = Disabled
When Sequence On Noread is enabled, the master symbol will increment by one if
either a match or a noread occurs.
If both Sequence On Noread and Sequence On Mismatch are enabled, the master
symbol will increment in every event (match, mismatch, or noread).
In the special case, where the expected number is missed by only 1, the sequence skips
over the next number in order to resume the sequence. For example, a 002 is expected
but a 003 is read. The next master symbol will be 004.See the tables under
On Mismatch" on page 6-8
character,sequence on
1 = Enabled
for examples of this.
Chapter 6 Matchcode
noread,sequence on mis-
MS-9 Scanner User's Manual
"Sequence
6-7