Keithley 2001 Operator's Manual page 68

Multimeter
Hide thumbs Also See for 2001:
Table of Contents

Advertisement

Front Panel Operation
+00.00100 Ω
PREV
DISPLAY
NEXT
POWER
Note: Source current flows from the
cáÖìêÉ=PJNQ
Two-wire resistance measurements
6. Observe the display. If the "Overflow" message is
shown, select a higher range until a normal reading is
displayed. Always use the lowest possible range for the
best resolution.
7. Take a reading from the display.
4-wire resistance measurements
The Model 2001 can make 4-wire resistance measurements
between 1µ and 210k. Assuming "bench reset"
conditions (see
paragraph
follows:
1. Connect test leads to the INPUT HI and LO and SENSE
4 WIRE HI and LO terminals of the Model 2001. Rec-
ommended Kelvin test probes include the Keithley
Models 5805 and 5806. Either the front or rear inputs
can be used; place the INPUTS button in the appropriate
position.
2. Select the 4 function.
3. Select a range consistent with the expected resistance.
For automatic range selection, press the AUTO key. The
AUTO annunciator denotes whether auto-ranging is en-
abled.
4. Enable offset compensation if needed (refer to the pro-
cedure later in this paragraph).
If offset compensation is not being used,
the 20 and 200 ranges require zero
correction in order to achieve the best
accuracy. The zero correction procedure is
located in a following paragraph.
5. Connect the test leads to the resistance as shown in
Figure
3-15.
3-36
Model 2001
Range: 20Ω
2001 MULTIMETER
Ω2
Ω4
DCV
ACV
DCI
ACI
FREQ
TEMP
REL
TRIG
STORE RECALL
FILTER
MATH
INFO
LOCAL
CHAN
SCAN
CONFIG
MENU
EXIT
ENTER
INPUT HI to INPUT LO terminals.
3.12.1), the basic procedure is as
Shielded
Cable
SENSE
INPUT
Ω 4 WIRE
HI
350V
1100V
PEAK
PEAK
LO
500V
INPUTS
PEAK
F
R
RANGE
AUTO
FRONT/REAR
2A 250V
AMPS
RANGE
CAL
6. Observe the display. If the "Overflow" message is
shown, select a higher range until a normal reading is
displayed. Always use the lowest possible range for the
best resolution.
7. Take a reading from the display.
Zeroing
The term "when properly zeroed" means that you must
establish a baseline for subsequent measurements on that
range. The 20 and 200 resistance ranges require zero
correction to correct for thermal offsets. This procedure
should be performed whenever the ambient temperature
changes. To zero (rel) the Model 2001, use the following
procedure:
1. Disable rel, if presently enabled, by pressing the REL
key. The REL annunciator will turn off.
2. Select the desired function (2 or 4) and range.
3. Connect the test leads to INPUT HI and LO (and
SENSE 4 WIRE if the 4 function is selected) of the
Model 2001 and short them together. Noise and thermal
offsets may require a few moments to stabilize.
4. Press the REL key. The display will read zero.
5. Remove the short and connect the test leads to the resis-
tance to be measured.
Optional shield
Resistance
Under Test
2001-900-01 Rev. K/ August 2010

Advertisement

Table of Contents
loading

Table of Contents