Keithley 2001 Operator's Manual page 406

Multimeter
Hide thumbs Also See for 2001:
Table of Contents

Advertisement

SCPI Command Subsystems
q~ÄäÉ=`JT=E`çåíáåìÉÇF
SENSe command summary
[:SENSe[1]]
:CURRent AC
:AVERage
:ADVanced
:NTOLerance <n>
:NTOLerance?
[:STATe] <b>
[:STATe]?
[:STATe] <b>
[:STATe]?
:AUTO <b>
:AUTO ONCE
:AUTO?
:DETector
[:FUNCtion] <name>
[:FUNCtion]?
:CURRent[:DC]
:APERture <n>
:AUTO <b>
:AUTO ONCE
:AUTO?
:APERture?
:NPLCycles <n>
:AUTO <b>
:AUTO ONCE
:AUTO?
:NPLCycles?
:RANGe
[:UPPer] <n>
[:UPPer]?
:AUTO <b>
:AUTO ONCE
:ULIMit <n>
:ULIMit?
:LLIMit <n>
:LLIMit?
:AUTO?
:REFerence <n>
:STATe <b>
:STATe?
:ACQuire
:REFerence?
:DIGits <n>
:AUTO <b>
:AUTO ONCE
:AUTO?
:DIGits?
C-8
Path to configure and control advanced filter.
Specify noise tolerance level (0 to 100 percent).
Query noise tolerance level.
Enable (1 or ON) or disable (0 or OFF) advanced filter.
Query state of advanced filter.
Enable (1 or ON) or disable (0 or OFF) filter.
Query state of digital filter.
Enable (1 or ON) or disable (0 or OFF) auto filter.
Enable and then disable auto filter.
Query auto filter.
Path to select function:
Select type of AC measurement: RMS, AVERage.
Query detector function.
Path to configure DC current.
Specify integration rate in seconds (166.67e-6 to 200e-3).
Enable (1 or ON) or disable (0 or OFF) auto aperture.
Enable and then disable auto aperture.
Query auto aperture (0 or 1).
Query aperture (integration rate).
Specify integration rate (number of line cycles; 0.01 to 10).
Enable (1 or ON) or disable (0 or OFF) auto NPLC.
Enable and then disable auto NPLC.
Query auto line cycle integration (0 or 1).
Query line cycle integration rate.
Path to configure measurement range:
Select range (0 to 2.1).
Query range.
Enable (1 or ON) or disable (0 or OFF) auto range.
Set range based on present input signal.
Specify upper limit for auto range (0 to 2.1).
Query upper limit.
Specify lower limit for auto range (0 to 2.1).
Query lower limit.
Query auto range.
Specify reference (REL) value (-2.1 to +2.1).
Enable (1 or ON) or disable (0 or OFF) REL.
Query state of REL (0 or 1).
Use input signal as reference.
Query reference (REL) value.
Specify measurement resolution (4 to 9).
Enable (1 or ON) or disable (0 or OFF) auto resolution.
Enable and then disable auto resolution.
Query auto resolution (0 or 1).
Query resolution.
2001-900-01 Rev. K / August 2010