Output Subsystem - Keithley 2001 Operator's Manual

Multimeter
Hide thumbs Also See for 2001:
Table of Contents

Advertisement

IEEE-488 Reference
Parameters
Query
Description
Programming example
4.17

:OUTPut subsystem

q~ÄäÉ=QJNM
OUTPut command summary
Command
:OUTPut
:TTL[1]
:LSENse <name>
:LSENse?
:TTL2
:LSENse <name>
:LSENse?
:TTL3
:LSENse <name>
:LSENse?
:TTL4
:LSENse <name>
:LSENse?
Notes:
1. Brackets [ ] are used to denote optional character sets. These optional characters do not have to be included in the pro-
gram message. Do not use brackets [ ] in the program message.
2. Angle brackets < > are used to indicate parameter type. Do not use angle brackets < > in the program message.
3. Upper case characters indicate the short-form version for each command word.
Parameters
4-88
:EXPonent <name>
:FORMat:EXPonent <name>
<name> = NORMal
= HPRecision High-precision format
:EXPonent?
Query exponent format
This command defines the exponent format. The default setting is NORMal, in which case the
returned value is formatted according to the engineering units of the present range (for example,
0.0012341E+03 on the 2k, 20k, or 200k range). Choosing HPRecision selects the highest preci-
sion possible, so the above example would be returned as 1.2341234E+00 with three extra digits
of precision.
OUTPUT 716; ":form:exp hpr"
The OUTPut subsystem is used to set polarities for the digital output port. Commands in this
subsystem are summarized in
Description
Path to set polarity of digital output line 1:
Select polarity; active high (AHIGH) or active low (ALOW).
Query polarity of line 1.
Path to set polarity of digital output line 2:
Select polarity; active high (AHIGH) or active low (ALOW).
Query polarity of line 2.
Path to set polarity of digital output line 3:
Select polarity; active high (AHIGH) or active low (ALOW).
Query polarity of line 3.
Path to set polarity of digital output line 4:
Select polarity; active high (AHIGH) or active low (ALOW).
Query polarity of line 4.
:LSENse <name>
:OUTPut:TTL[1]:LSENse <name>
:OUTPut:TTL2:LSENse <name>
:OUTPut:TTL3:LSENse <name>
:OUTPut:TTL4:LSENse <name>
<name> = AHIGh
Set exponent format.
Normal format
Table
4-10.
Set polarity of line #1
Set polarity of line #2
Set polarity of line #3
Set polarity of line #4
Set polarity of specified line to active high.
2001-900-01 Rev. K / August 2010

Advertisement

Table of Contents
loading

Table of Contents