Bit B10 Not used.
Bit B11, Buffer Pretriggered (BPT) Set bit indicates that the trace buffer pretrigger event has oc-
curred (PTR), or the operations associated with the pretrigger have been completed (NTR).
Bits B12 through B15 Not used.
Questionable Event Register:
Bits B0 through B3 Not used.
Bit B4, Temperature Summary (Temp) Set bit indicates that an invalid reference junction measure-
ment has occurred for thermocouple temperature measurements (PTR), or a subsequent valid reference
junction measurement has occurred (NTR).
Bits B5, B6 and B7 Not used.
Bit B8, Calibration Summary (Cal) Set bit indicates that an invalid calibration constant was detected
during the power-up sequence (PTR). The instrument will instead use a default calibration constant. This
error will clear after successful calibration of the instrument.
Bits B9 through B13 Not used.
Bit B14, Command Warning (Warn) Set bit indicates that a Signal Oriented Measurement Com-
mand parameter has been ignored (PTR), or a subsequent Signal Oriented Measurement Command has
executed successfully (NTR).
Note: Whenever a questionable event occurs, the ERR annunciator will turn on. The annunciator will turn
off when the questionable event clears.
2001-900-01 Rev. K / August 2010
Bit Position
B15 - B12
Event
Decimal Weighting
Value
Value : 1 = Measurement Event Set
0 = Measurement Event Cleared
cáÖìêÉ=QJOO
Measurement Event Register
B11
B10
B9
B8
B7
B6
BPT
BFL
BHF BAV
2048
512
256
128
(2 )
11
(2 )
9
(2 )
8
(2 )
7
0/1
0/1
0/1
0/1
Events : BPT = Buffer Pretriggered
IEEE-488 Reference
B5
B4
B3
B2
B1
B0
RAV
HL2
LL2
HL1
LL1
ROF
32
2
16
8
4
1
(2 )
5
(2 )
4
(2 )
3
(2 )
2
(2 )
1
(2 )
0
0/1
0/1
0/1
0/1
0/1
0/1
BFL = Buffer Full
BHF = Buffer Half Full
BAV = Buffer Available
RAV = Reading Available
HL2 = High Limit 2
LL2 = Low Limit 2
HL1 = High Limit 1
LL1 = Low Limit 1
ROF = Reading Overflow
4-167