Keithley 2001 Operator's Manual page 236

Multimeter
Hide thumbs Also See for 2001:
Table of Contents

Advertisement

IEEE-488 Reference
Defaults
Query
Description
Programming example
Parameters
Formats
Defaults
4-78
where; Y =
number of characters in message:
Up to 20 for top display.
Up to 32 for bottom display.
X =
number of digits that make up Y (1 or 2).
Power-up
*RST
:SYSTem:PRESet
:DATA?
Query the defined text message
Short-form formats:
:disp:text:data?
:disp:wind2:text:data?
Response message:
User defined text message
These commands define the text messages that you wish to display. The message can be as long
as 20 characters for the top display, and up to 32 characters for the bottom display. A space is
counted as a character. Excess message characters (over 20 and 32 respectively) will cause error.
A string message must be enclosed in single quotes or double quotes. Both of the following
string messages are valid:
'My dog has fleas' or "My dog has fleas"
An indefinite block message must be the only command in the program message or the last
command in the program message. If you include a command after an indefinite block message
(on the same line), it will be treated as part of the message and will be displayed instead of ex-
ecuted.
10 OUTPUT 716; ":disp:text:data 'Keithley Model 2001'; data?"
20 ENTER 716; A$
30 PRINT A$
40 END
Line 10
Defines the user defined message for the top display, and then queries the defined
message.
Line 20
Addresses the Model 2001 to talk.
Line 30
Displays (on the CRT) the user defined message.
:STATe <b>
:DISPlay[:WINDow[1]]:TEXT:STATe <b>
:DISPlay:WINDow2:TEXT:STATe <b>
<b>
= 0 or OFF
= 1 or ON
:disp:text:stat <b>
:disp:wind2:text:stat <b>
Power-up
Null string
No effect
No effect
Control (on/off) message; top display
Control (on/off) message; bottom display
Disable text message for specified display
Enable text message for specified display
OFF
2001-900-01 Rev. K / August 2010