IEEE-488 Reference
Parameters
Formats
Defaults
Query
Description
Programming example
4-142
[:STATe] <b>
[:SENSe[1]]:CURRent:AC:AVERage[:STATe] <b>Control filter for ACI
[:SENSe[1]]:CURRent[:DC]:AVERage[:STATe] <b>Control filter for DCI
[:SENSe[1]]:VOLTage:AC:AVERage[:STATe] <b>Control filter for ACV
[:SENSe[1]]:VOLTage[:DC]:AVERage[:STATe] <b>Control filter for DCV
[:SENSe[1]]:RESistance:AVERage[:STATe] <b>Control filter for 2
[:SENSe[1]]:FRESistance:AVERage[:STATe] <b>Control filter for 4
[:SENSe[1]]:TEMPerature:AVERage[:STATe] <b>Control filter for TEMP
<b>
= 0 or OFF
= 1 or ON
:curr:ac:aver <b>
:curr[:dc]:aver <b>
:volt:ac:aver <b>:volt[:dc]:aver <b>
:res:aver <b>:fres:aver <b>
:temp:aver <b>
Power-up
*RST
:SYSTem:PRESet
:STATe]?
Query state of digital filter
Short-form format:
:curr:ac:aver?
:volt:ac:aver?
:res:aver?
:temp:aver?
Response message:
0 (off) or 1 (on)
These commands are used to enable or disable the digital averaging filter for the specified func-
tion. When enabled, readings will be filtered according to how the filter is configured.
10 OUTPUT 716; ":curr:ac:aver on; aver?"
20 ENTER 716; A$
30 PRINT A$
40 END
Line 10
Enables the filter, and then queries the state of the filter.
Line 20
Addresses the Model 2001 to talk.
Line 30
Displays the state of the filter (1; on).
Disable the digital filter
Enable the digital filter
Saved power-on setup
OFF (all functions)
OFF (ACI, ACV and TEMP)
ON (DCI, DCV, 2 and 4)
:curr[:dc]:aver?
:volt[:dc]:aver?
:fres:aver?
2001-900-01 Rev. K / August 2010