Keithley 2001 Operator's Manual page 379

Multimeter
Hide thumbs Also See for 2001:
Table of Contents

Advertisement

Programming example
Parameters
Formats
Defaults
Query
Description
2001-900-01 Rev. K / August 2010
10 OUTPUT 716; ":trig:tcon:dir sour; dir?"
20 ENTER 716; A$
30 PRINT A$
40 END
Line 10
Enables the Source Bypass for the measure layer and then queries the state of the Source By-
pass.
Line 20
Addresses the Model 2001 to talk.
Line 30
Displays the state of the Source Bypass (SOUR; enabled).
:ASYNchronous commands
:ILINe <NRf>
:ARM[:SEQuence[1]][:LAYer[1]]:TCONfigure:ASYNchronous:ILINe <NRf>Select input trigger line; arm layer
:ARM[:SEQuence[1]]:LAYer2:TCONfigure:ASYNchronous:ILINe <NRf>Select input trigger line; scan layer
:TRIGger[:SEQuence[1]]:TCONfigure:ASYNchronous:ILINe <NRf>Select input trigger line; measure layer
<NRf>
= 1
= 2
= 3
= 4
= 5
= 6
:arm:tcon:asyn:ilin <NRf>
:arm:lay2:tcon:asyn:ilin <NRf>
:trig:tcon:asyn:ilin <NRf>
Power-up
*RST
:SYSTem:PRESet
:ILINe?
Query programmed input line #
Short-form format:
:arm:tcon:asyn:ilin?
:arm:lay2:tcon:asyn:ilin?
:trig:tcon:asyn:ilin?
Response message:
1 to 6
These commands are used to select an input line for the asynchronous Trigger Link of the specified layer.
Keep in mind that asynchronous Trigger Link input and output (see :OLINe) cannot share the same line.
If you assign the input to a line that is already being used for the output, an error will occur and the com-
mand will be ignored.
Line #1
Line #2
Line #3
Line #4
Line #5
Line #6
Saved power-on setup
2 (all layers)
2 (all layers)
IEEE-488 Reference
4-221

Advertisement