Keithley 2001 Operator's Manual page 458

Multimeter
Hide thumbs Also See for 2001:
Table of Contents

Advertisement

SCPI Conformance Information
Command
:DISPlay
[:WINDow[1]]
:DATA?
:WINDow2
:DATA?
:CNDisplay
:SMESsage <b>
:SMESsage?
:FORMat
:ELEments <item list>
:ELEments?
:OUTPut
:TTL[1]
:LSENse <name>
:LSENse?
:TTL2
:LSENse <name>
:LSENse?
:OUTPut
:TTL3
:LSENse <name>
:LSENse?
:TTL4
:LSENse <name>
:LSENse?
:ROUTe
:SCAN
[:INTernal]
:FUNCtion? <list>, <name>
:FUNCtion? <list>
:EXTernal <list>
:EXTernal?
:FUNCtion? <list>, <name>
:FUNCtion? <list>
:RATio
:RCHannel <list>
:RCHannel?
:MCHannel <list>
:MCHannel?
:DELTa
:RCHannel <list>
:RCHannel?
:MCHannel <list>
:MCHannel?
:LSELect <name>
:LSELect?
I-10
Description
Path to locate message to top display.
Query data on top portion of display.
Path to locate message to bottom display.
Query data on bottom portion of display.
Clear NEXT (or PREV) display messages and cancels associated operations.
Enable (1 or ON) or disable (0 or OFF) status message mode.
Query status message mode (0 or 1).
Specify data elements: READing, CHANnel, RNUMber, UNITs, TIMEstamp and
STATus.
Query data elements.
Path to set polarity of digital output line 1:
Select polarity; active (AHIGH) or active (ALOW).
Query polarity of line 1.
Path to set polarity of digital output line 2:
Select polarity; active (AHIGH) or active (ALOW).
Query polarity of line 2.
Path to set polarity of digital output line 3:
Select polarity; active (AHIGH) or active (ALOW).
Query polarity of line 3.
Path to set polarity of digital output line 4:
Select polarity; active (AHIGH) or active (ALOW).
Query polarity of line 4.
Command path to configure and control scan operations:
Specify an internal scan list (2 to 10 channels).
Assign measurement function to specified channels.
Query function for each specified channel.
Specify an external scan list (2 to 80 channels).
Query the external scan list.
Assign measurement function to specified channels.
Query function for each specified channel.
Command path to configure ratio calculation:
Specify reference channel (1 to 10).
Query reference channel.
Specify measure channel (1 to 10).
Query measure channel.
Command path to configure delta calculation:
Specify reference channel (1 to 10).
Query reference channel.
Specify measure channel (1 to 10).
Query measure channel.
Select scan operation: INTernal, EXTernal, RATio, DELTa, NONE.
Query scan operation.
2001-900-01 Rev. K / August 2010