Getting Started
130 OUTPUT 716;":trac:clear"
140 OUTPUT 716;":trac:feed calc"
150 OUTPUT 716;":trac:poin 100"
160 OUTPUT 716;":init"
170 WAIT 2
180 OUTPUT 716;":trac:data?"
190 ENTER 716;A$
200 PRINT A$
Line 100 Return Model 2001 to default configuration.
Line 110 Change acquisition method to burst; put unit in
idle.
Line 120 Specify data elements (reading, reading number,
units, and status).
Line 130 Clear all stored readings.
Line 140 Perform any math before storing readings.
Line 150 Set buffer count to 10 readings.
Line 160 Take unit out of idle; perform reading burst;
return to idle.
Line 170 Wait for burst to complete.
Line 180 Request readings (default is readings only, ASCII
format).
Line 190 Address to talk.
Line 200 Display readings.
2-18
2001-900-01 Rev. K / August 2010