Keithley 2001 Operator's Manual page 106

Multimeter
Hide thumbs Also See for 2001:
Table of Contents

Advertisement

Front Panel Operation
COM OUT
230
Voltage
Source
cáÖìêÉ=PJPP
DUT test system (asynchronous example #2)
7001
Switch
System
230
Voltage
External Trigger
Source
cáÖìêÉ=PJPQ
Trigger Link connections (asynchronous example #2)
For this example, the Model 230 is programmed for External
Triggering and is set to source the first voltage level. The
Models 2001 and 7001 are configured as follows:
Model 2001:
Idle state:
Bench reset = :INIT:CONT
Arm layer:
Arm source = Immediate*
Arm count = 1*
Arm trigger control = Acceptor*
3-74
DUT
1
#1
DUT
2
#2
DUT
10
#10
IN
OUT
Trigger
Link
IN
OUT
ON*
OUTPUT
Card 1
7011 MUX Card
Trigger
Link Cables
(8501)
8502
Trigger Link
Adapter
1
2 3
4
5
6
BNC to BNC
Cables
(7501)
Scan layer:
Scan source = Immediate*
Scan count = Infinite*
Scan trigger control = Acceptor*
Measure layer:
Measure source = TrigLink
Trigger link mode = Asynchronous
Input line = #3
Output line = #4
Measure count = 20
Measure trigger control = Acceptor*
Input HI
Input LO
2001 Multimeter
IN OUT
Trigger
Link
2001 Multimeter
2001-900-01 Rev. K/ August 2010