Keithley 2001 Operator's Manual page 410

Multimeter
Hide thumbs Also See for 2001:
Table of Contents

Advertisement

SCPI Command Subsystems
q~ÄäÉ=`JT=E`çåíáåìÉÇF
SENSe command summary
[:SENS[1]]
:RESistance
:RANGe
:AUTO ONCE
:LLIMit <n>
:LLIMit?
:AUTO?
:REFerence <n>
:STATe <b>
:STATe?
:ACQuire
:REFerence?
:DIGits <n>
:AUTO <b>
:AUTO ONCE
:AUTO?
:DIGits?
:AVERage
:TCONtrol <name>
:TCONtrol?
:COUNt <n>
:COUNt?
:ADVanced
:NTOLerance <n>
:NTOLerance?
[:STATe] <b>
[:STATe]?
[:STATe] <b>
[:STATe]?
:AUTO <b>
:AUTO ONCE
:AUTO?
:OCOMpensated <b>
:OCOMpensated?
:FRESistance
:APERture <n>
:AUTO <b>
:AUTO ONCE
:AUTO?
:APERture?
:NPLCycles <n>
:AUTO <b>
:AUTO ONCE
:AUTO?
:NPLCycles?
:RANGe
[:UPPer] <n>
[:UPPer]?
:AUTO <b>
C-12
Specify lower limit for auto range (0 to 1.05e9).
Query lower limit.
Query auto range.
Specify reference (REL) value (0 to 1.05e9).
Enable (1 or ON) or disable (0 or OFF) REL.
Query state of REL (0 or 1).
Use input signal as reference.
Query reference (REL) value.
Specify measurement resolution (4 to 9).
Enable (1 or ON) or disable (0 or OFF) auto resolution.
Enable and then disable auto resolution.
Query auto resolution (0 or 1).
Query resolution.
Path to configure and control the filter.
Select filter type: MOVing, REPeat.
Query filter type.
Specify filter count (1 to 100).
Query filter count.
Path to configure and control advanced filter.
Specify noise tolerance level (0 to 100 percent).
Query noise tolerance level.
Enable (1 or ON) or disable (0 or OFF) advanced filter.
Query state of advanced filter.
Enable (1 or ON) or disable (0 or OFF) filter.
Query state of digital filter.
Enable (1 or ON) or disable (0 or OFF) auto filter.
Enable and then disable auto filter.
Query auto filter.
Enable (1 or ON) or disable (0 or OFF) Offset Compensation.
Query Offset Compensation.
Path to configure four-wire resistance:
Specify integration rate in seconds (166.67e-6 to 200e-3).
Enable (1 or ON) or disable (0 or OFF) auto aperture.
Enable and then disable auto aperture.
Query auto aperture (0 or 1).
Query aperture (integration rate).
Specify integration rate in number of line cycles (0.01 to 10).
Enable (1 or ON) or disable (0 or OFF) auto NPLC.
Enable and then disable auto NPLC.
Query auto line cycle integration (0 or 1).
Query line cycle integration rate.
Path to configure measurement range:
Select range (0 to 2.1e5).
Query range.
Enable (1 or ON) or disable (0 or OFF) auto range.
2001-900-01 Rev. K / August 2010

Advertisement

Table of Contents
loading

Table of Contents