Keithley 2001 Operator's Manual page 110

Multimeter
Hide thumbs Also See for 2001:
Table of Contents

Advertisement

Front Panel Operation
Scan source = Immediate*
Scan count = Infinite*
Scan trigger control = Acceptor*
Measure layer:
Measure source = TrigLink
Trigger link mode = Semi-synchronous
Semi-sync line = #1*
Measure count = 10
Measure trigger control = Acceptor*
* Indicates that the setting is the BENCH RESET (and
factory) default condition.
Model 7001:
Idle state:
Reset = :INIT:CONT OFF*
Scan list = 1!1-1!10,
Arm layer:
Arm spacing = Immediate*
Arm count = 1*
Arm trigger control = Acceptor*
Scan layer:
7001 Switch System
cáÖìêÉ=PJPU
Trigger Link connections (semi-synchronous example)
A
The BENCH RESET condition arms the Model 2001
and places multimeter operation at point A in the flowchart,
where it is waiting for a Trigger Link trigger. Note that since
both the arm layer and scan layer are programmed for
Immediate Source, operation immediately drops down to the
measure layer at point A.
B
Pressing STEP takes the Model 7001 out of the idle state
and places operation at point B in the flowchart. Since both
the arm layer and scan layers are programmed for Immediate
Spacing, operation drops down to the Channel Layer at point
B.
3-78
Scan spacing = Immediate*
Number of scans = 1
Scan trigger control = Acceptor*
Channel layer:
Channel spacing = TrigLink
Trigger link mode = Semi-synchronous
Semi-sync line = #1
Number of channels = Use Scanlist length*
Channel trigger control = Source*
* Indicates that the setting is the RESET (and factory)
default condition.
To run the test and store the readings in the Model 2001,
press STORE on the multimeter, enter the desired number of
readings (ten), and press ENTER. The Model 2001 waits
(with the asterisk annunciator lit) for a Trigger Link trigger
from the Model 7001. Press STEP on the Model 7001 to start
the scan.
The following explanation on operation is referenced to the
operation model shown in
IN
Line #1
OUT
Trigger
Link
Trigger Link
Cable
(8501)
C
does not wait at point B for a trigger. Instead, it bypasses
"Wait for Trigger Link Trigger" and closes the first channel
(point C). Note that the Bypass is in effect only on the first
pass through the model.
D
Trigger Link trigger line (point D). Since the instrument is
programmed to scan ten channels, operation loops back up to
point B, where it waits for an input trigger. Note that Bypass
is no longer in effect.
E
point A waiting for a trigger. When the trigger line is pulled
Figure
2001 Multimeter
Since Channel Trigger Source is set to Source, the scan
After the relay settles, the Model 7001 pulls down the
F
and
Remember that the Model 2001 operation is at
2001-900-01 Rev. K/ August 2010
3-39.
IN OUT
Trigger
Link