Sprtd Commands - Keithley 2001 Operator's Manual

Multimeter
Hide thumbs Also See for 2001:
Table of Contents

Advertisement

Programming example

4.19.20 :SPRTD commands

Parameters
Query
Description
2001-900-01 Rev. K / August 2010
10 OUTPUT 716; ":temp:rtd:rzer 200; rzer?"
20 ENTER 716; A$
30 PRINT A$
40 END
Line 10
Two commands in this program message; the first changes the resistance value to 200, and
the second queries the resistance value.
Line 20
Addresses the Model 2001 to talk.
Line 30
Displays the resistance value (200).
[:SENSe[1]]:TEMPerature:SPRTD:RZERo <NRf>
[:SENSe[1]]:TEMPerature:SPRTD:A4 <NRf>
[:SENSe[1]]:TEMPerature:SPRTD:B4 <NRf>
[:SENSe[1]]:TEMPerature:SPRTD:A7 <NRf>
[:SENSe[1]]:TEMPerature:SPRTD:B7 <NRf>
[:SENSe[1]]:TEMPerature:SPRTD:C7 <NRf>
:RZERo <NRf>Set RTD 0°C resistance value
:A4 <NRf>
:B4 <NRf>
:A7 <NRf>
:B7 <NRf>
:C7 <NRf>
:RZERo?
Query RTD 0°C resistance value
:A4?
Query A4 coefficient
:B4?
Query B4 coefficient
:A7?
Query A7 coefficient
:B7?
Query B7 coefficient
:C7?
Query C7 coefficient
These commands program the various parameters for SPRTD (Standard Platinum Resistance Thermom-
eters). A single SPRTD, however, usually cannot be used to cover the entire range, so the temperature
range is broken up into several subranges. An SPRTD is supplied with a certificate of calibration that lists
the calibration constants and the temperature range supported. In most cases, the supplied coefficients
Set A4 coefficient
Set B4 coefficient
Set A7 coefficient
Set B7 coefficient
Set C7 coefficient
IEEE-488 Reference