Keithley 2001 Operator's Manual page 336

Multimeter
Hide thumbs Also See for 2001:
Table of Contents

Advertisement

IEEE-488 Reference
4-178
Effects of positive transitions on the Questionable Event Register:
Questionable event
Temperature summary
Calibration summary
Command warning
Bit Position
Event
Decimal Weighting
Value
Value : 1 = Enable Positive Transition
A) Positive Transition (PTR) Register
Bit Position
Event
Decimal Weighting
Value
Value : 1 = Enable Negative Transition
B) Negative Transition (NTR) Register
cáÖìêÉ=QJPR
Questionable Transition Filter
Positive transition effect on
Questionable Event Register
Sets B4 when an invalid reference junction measurement has oc-
curred.
Sets B8 when an invalid calibration constant is detected on power-up.
Sets B14 when a Signal Oriented Measurement Command parameter
is ignored.
B15
B14
B13 - B9
B8
Warn
Cal
25
16384
6
(2 )
(2 )
14
0
0/1
0/1
0 = Disable Positive Transition
B15
B14
B13 - B9
Warn
Cal
16384
256
(2 )
14
(2 )
0
0/1
0/1
0 = Disable Negative Transition
B7 - B5
B4
B3 - B0
Temp
16
8
(2 )
4
0/1
B8
B7 - B5
B4
B3 - B0
Temp
16
8
(2 )
4
0/1
2001-900-01 Rev. K / August 2010