Keithley 2001 Operator's Manual page 347

Multimeter
Hide thumbs Also See for 2001:
Table of Contents

Advertisement

Programming example
Parameter
Format
Defaults
Query
Description
Programming example
2001-900-01 Rev. K / August 2010
Note: To disable all messages from entering the Error Queue, send the following command:
:stat:que:enab ()
10 OUTPUT 716; ":stat:que:enab (0:-999); enab?"
20 ENTER 716; A$
30 PRINT A$
40 END
Line 10
Enables all SCPI defined messages.
Line 20
Addresses the Model 2001 to talk.
Line 30
Displays the messages (0:-999).
DISable <list>
:STATus:QUEue:DISable <list>
<list> = (numlist)
where; numlist is a specified list of messages that you wish to disable for the Error Queue.
:stat:que:dis (numlist)
Power-up
*CLS
:STATus:PRESet
:DISable?
Query list of disabled messages
Short-form format:
:stat:que:dis?
Response message:
numlist
On power-up, all error messages are enabled and will go into the Error Queue as they occur. Status mes-
sages are not enabled and will not go into the queue. This command is used to specify which messages
you want disabled. Disabled messages are prevented from going into the Error Queue.
Messages are specified by numbers (see
pressing a message numlist:
Numlist = -110
= -110, -140, -222
= -110:-222
= -110:-222, -230
10 OUTPUT 716; ":stat:que:dis (-140:-150); dis?"
20 ENTER 716; A$
30 PRINT A$
40 END
Line 10
Disables messages -140 through -150.
Line 20
Addresses the Model 2001 to talk.
Line 30
Displays the disabled messages (-140:-150).
Disable messages for Error Queue
Clears list
No effect
No effect
Table
3-4). The following examples show various forms for ex-
Single message.
Messages separated by commas.
Range of messages (-100 through -222).
Range entry and single entry separated by a comma.
IEEE-488 Reference
4-189