Getting Started
q~ÄäÉ=OJQ
Multiple displays for recalled readings
+00.00000 VDC
Rdg#+00000
sec
NEXT 
+00.00000 VDC
MAX=+0.000000e+00 at RDG# +00000
NEXT 
+00.00000 VDC
MIN=+0.000000e+00 at RDG# +00000
NEXT 
+00.00000 VDC
AVG=+0.0000e+00 SDEV=+0.0000e+00
NEXT 
PRINT BUFFER DATA
Press ENTER to start printing.
NEXT 
Note: Press the NEXT and PREV DISPLAY keys to scroll through the multiple displays (with wraparound.)
Step 4
Configure for burst of high speed readings
There are two data acquisition modes, normal and burst. The
burst reading rate is 2000 readings/sec into the buffer (plus
post-processing time of the raw readings, typically 2msec/
reading). To achieve this speed, certain tradeoffs are made
with flexibility, as listed in
1. To continue the reading storage example with burst
mode selected, display the CONFIG DATA STORE
menu by pressing the CONFIG key, then the STORE
key:
CONFIG DATA STORE
BURST-MODE DATA-GROUP CONTROL
CLEAR-ALL COUNT FEED
2. Select BURST-MODE and press ENTER. After a
momentary message about clearing the buffer, the
display shows:
BURST MODE
OFF ON
3. Select ON and press ENTER. (Certain instrument
parameters are saved when burst is enabled, and are
restored when burst is disabled. See
further details on burst mode.)
2-12
@Time=+000.000000
PREV
PREV
PREV
PREV
PREV
Table
2-5.
paragraph 3.8.1
Normal stored reading.
Reading number and time-stamp.
Maximum value of stored readings.
Minimum value of stored readings.
Average and standard deviation.
Prompt to dump readings to a printer.
q~ÄäÉ=OJR
Burst mode availability
DC voltage
AC voltage
DC current
AC current
2-wire resistance
4-wire resistance
Frequency
Temperature
Range
Autozero
Speed
Filter
Resolution
Multiple displays
Data grouping
Delay
Notes:
1. The measurement display is not updated during burst.
2. A math operation slows post-processing time.
for
Normal (no peak spikes)
Normal RMS or average
Normal (no in-circuit)
RMS or average
Normal (no offset compensation)
Not available
Not available
Not available
Fixed
Disabled
FAST (0.01 PLC)
Repeat acts like moving.
4.5 digits
Disabled
Compact (reading and reading
number)
As set.