Rtd Commands - Keithley 2001 Operator's Manual

Multimeter
Hide thumbs Also See for 2001:
Table of Contents

Advertisement

IEEE-488 Reference

4.19.19 :RTD commands

Parameters
Format
Defaults
Query
Description
Programming example
Parameter
4-152
:TYPE <name>
[:SENSe[1]]:TEMPerature:RTD:TYPE <name>Set RTD parameters
<name> = PT100
= D100
= F100
= PT385
= PT3916
= USER
= SPRTD
:temp:rtd:type <name>
Power-up
*RST
:SYSTem:PRESet
:TYPE?
Query RTD type
Short-form format:
:temp:rtd:type?
Response message:
PT100, D100, F100, PT385, PT3916, USER or SPRTD
This command is used to select the RTD standard and other related factors. When PT100, D100,
F100, PT385 or PT3916 is selected, the instrument defaults to the following RTD factors:
Type
Standard
PT100
ITS-90
D100
ITS-90
F100
ITS-90
PT385
IPTS-68
PT3916
IPTS-68
Changing alpha (see :ALPHa), beta (see :BETA), delta (see :DELTa) or  at 0°C (see :RZERo)
automatically changes the type to USER . Therefore, once you program the appropriate factors
for your RTD, you can later recall that setup by using the :TYPE USER command.
With the SPRTD type selected, refer to
for standard platinum resistance thermometers.
10 OUTPUT 716; ":temp:rtd:type pt385; type?"
20 ENTER 716; A$
30 PRINT A$
40 END
Line 10
Two commands in this program message; the first programs the RTD factors to the
PT385 defaults, and the second queries the Type.
Line 20
Addresses the Model 2001 to talk.
Line 30
Displays Type (PT385).
:ALPHa <NRf>
[:SENSe[1]]:TEMPerature:RTD:ALPHa <NRf>Specify Alpha value
<NRf> = 0 to 0.01
Selects default parameters for the PT100 standard
Selects default parameters for the D100 standard
Selects default parameters for the F100 standard
Selects default parameters for the PT385 standard
Selects default parameters for the PT3916 standard
Selects user-defined parameters
Selects the SPRTD type.
Saved power-on setup
PT100
PT100
Alpha
Beta
0.003850
0.10863
0.003920
0.10630
0.003900
0.11000
0.003850
0.11100
0.003916
0.11600
paragraph 4.19.20
Specify RTD Alpha value
 at 0°C
Delta
1.49990
100
1.49710
100
1.49589
100
1.50700
100
1.50594
100
to program the various parameters
2001-900-01 Rev. K / August 2010

Advertisement

Table of Contents
loading

Table of Contents