Safety Requirements

READ BEFORE ATTEMPTING OPERATION OF THE INSTRUMENT
The Chandler Engineering Model 5265MGSA Gel Strength Analyzer is designed
with operator safety in mind. Any instrument that is capable of high temperatures
and pressures should always be operated with
NOTE: If this equipment is not used in a manner consistent with manufacturer's
specifications, the protection provided by the equipment may be
impaired.
WARNING: Read before attempting operation of instrument. This instrument is
capable of extremely high temperatures and pressures and must always
be operated with CAUTION. The instrument is designed for operator
safety. To ensure that safety it is essential to follow the instructions
outlined below.
To further ensure safety:
• Before attempting to operate the instrument, the operator should read and understand this
manual.
• If the instrument is not used in accordance with this manual, the safety of the instrument
may be impaired.
• Always position the instrument allowing easy access to the power cord.
• Post signs where the instrument is being operated to warn non-operating personnel that
high pressure, high temperature equipment is in use.
• Observe caution notes.
• Observe and follow the warning labels on the instrument.
• Never exceed the instrument's maximum temperature and pressure ratings.
• Always disconnect main power to the instrument before attempting any service or repair.
• A fire extinguisher, type 8 BC, should be located within 50 feet (15 meters) of
the instrument.
• Personal Protection Equipment (PPE) should be used when operating this instrument.
• It is the user's responsibility to ascertain the potential hazards of the samples under test.
Consult the appropriate MSDS and follow all PPE recommendations and dispose of
samples safely and appropriately.
• Never operated in hazardous atmospheres, applications, or with hazardous materials for
which the instrument is not designed.
• Provide adequate training for all personnel that will operate the instrument.
• Locate the instrument in a low traffic, well-ventilated area.
• This is a bench top device; place the associated instrument on a suitable level and stable
surface.