TITLE: PROC,CALIBRATION,5265MG
REF. NEXT ASSY: MODEL 5265MG GEL STRENGTH TESTER
9. Using the following table of masses and calibration values, individually enter the low and high Scaled
Values followed by selecting Add Sample.
 Low Calibration Value: 0 grams = 0 lbf/100ft
 High Calibration Value: 2000 grams = 530 lbf/100ft
10. Select OK to complete the calibration.
11. Optionally, additional masses are provided for use with determining the linearity of the calibration in
200 gram increments. Use the associated worksheet 5265MX-0009W.xlsx to determine and
document the load cell linearity.
12. Remove the calibrator assembly.
Part Number: 5265MX-0009
Page 2 of 3