TITLE: PROC, VALIDATION, MOTOR SPEED, 5270
REF. NEXT ASSY: MODEL 5265MG GEL STRENGTH TESTER
7. Select the 5265MG instrument to start a 5265MG test.
8. Select the Motor Speed Validation controller schedule.
9. Select "Start Test". Assign a file name when prompted.
10. Allow the test to run for approximately 2 minutes, followed by stopping the test.
11. Select the data tab below the plotted data and choose 2 points, 1 minute apart from the Motor Encoder column,
noting that the first point must be after the first non-zero Motor Encoder value.
Part Number: 5265MX-0122