SECTION 3 – OPERATING INSTRUCTIONS
3-5
Left Rear Panel
The left rear panel contains all the connections for the cables that connect the autoclave to the
processor.
Ultrasonic Connector
Not used on this instrument.
Thermocouple
The J-type thermocouple that is attached to the
top plug must be plugged into this connector
prior to the beginning of a test and any time
that it is necessary for the temperature
controller to operate. If the thermocouple is
not connected, the temperature controller will
display "SBr" and will not function.
Load Sensor or Expansion
This connection is used when the appropriate
add-on is purchased.
Communication Connections
Comm1, Comm2 and USB are serial
(Modbus) connections. Ethernet uses
ModbusTCP.
Power Connection
Requires an IEC cable to connect power. The
power required is 115/230 VAC, 50/60Hz,
12.5A.
NOTE: The Shrinkage connector is not
populated unless the appropriate add-on is
purchased.
Figure 7 – Left Rear Panel