10.3.6 Visual inspection
During the visual inspection, check all components for mechanical damage and leaks. Replace damaged
components immediately.
10.3.7 Leakage test
A leakage test is only possible if the QWIK-PURE
1. Fill the QWIK-PURE
2. Check all hose and other connections for leaks.
Symptoms
Hose connection leaking
Bayonet catch leaking
End cap leaking
iCS with tap water through the vent until the FRC performs a discharge operation.
®
iCS is completely filled with water.
®
Measure
•
Tighten the hose clamp.
•
Replace hardened hose and respective hose
clamps.
•
Check the fit of the seal and correct if necessary.
•
Check the seal for damage and replace if
necessary.
•
Tighten the bayonet fitting.
•
Check the seal for damage and replace if
necessary.
•
Check the fit of the seal and correct if necessary.
•
Check the seal for damage and replace if
necessary.
•
Tighten the end cap.