NOTICE
Filter cartridge insertion!
Use of incorrect filter cartridges or incorrect insertion of the filter cartridges can 
cause damage or leakage to the collector and the filter cartridges. 
Before inserting the filter cartridges, check to make sure that the filter 
cartridge is the right one for the product.
0 The color of the cap at the bottom of the filter cartridge must be identical 
to the color of the cap in the collector.
Insert the filter cartridges vertically and carefully into the collector.
Description / explanation
10. Insert the first filter cartridge into the mount 
on the foot with the bayonet mount facing the
measuring chamber outlet.
11. Turn the filter cartridge clockwise all the way. 
12. Align the connecting filter cartridge's connection 
with the connection on the measuring chamber
outlet.
13. Slide the bayonet mount over the connection
and turn it clockwise as far as it will go.
14. Insert the other filter cartridges into the holders 
and connect them together using the bayonet
catches.
15. Place the end caps on the last filter cartridge in 
each row and turn them clockwise all the way.