Check The Message System Of The Amalgam Separator - Dentsply Sirona Sinius Operating Instructions Manual

Hide thumbs Also See for Sinius:
Table of Contents

Advertisement

Dentsply Sirona
Operating Instructions Sinius
63  2 2  7 26 D3561
D3561.201.01.16.02    08.2018
5.5.7

Check the message system of the amalgam separator

An electronic control monitors the function of the amalgam separator. It
detects any mechanical blocking or failure of the drive motor. The error
is indicated by the Amalg display on the touchscreen and by an acoustic
signal.
The functionality of this message system must be checked at least once
per year.
Inform your service technician if the error occurs during regular operation.
In Germany: Document the test in the operation log of the amalgam
separator in the "Testing the function of display and message systems"
section.
ü All instruments are in their holders. The Start program is displayed
on the touchscreen.
1. Press and hold the Setup fixed key (> 2 s).
Ä The Setup dialog appears.
2. In the setup dialog, hold and press the Service key (> 2 s).
Ä The service dialog is shown.
3. Test the message system. Press and hold the Amalg key for an
extended period.
Ä The message system is working if a signal sounds as long as
the Amalg key is pressed.
Inform your service technician if the acoustic signal does not sound.
Press the Back key to exit the service dialog.
5 Care, cleaning and maintenance by the practice team
5.5 Components of the water unit
245

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sinius tsSinius cs