9.6 Accessories, User Replaceable Parts, and
Medicaments
9.6.1
Accessories
1.
Cavitron Select SPS Lavage Control Detachable Handpiece
Cable Assembly with Swivel
2. Steri-Mate Detachable Sterilizable Handpiece
3. Medically Approved Power Supply
4. AC Power Cord Set
5. Cavitron Ultrasonic Inserts
6. Dual Position Foot Control Assembly
7. Cavitron Select Reservoir Bottle
8. Cavitron Select Pump
For detailed information, contact your local Dentsply Sirona
Representative or authorized Dentsply Sirona Distributor.
9.6.2 User Replaceable Part Kits
1.
Cavitron Insert Replacement O-ring Kits, 12/packs
- Part Number 62351 (black) for plastic and soft grips
- Part Number 62605 (green) for metal grips
2. Handpiece Cable O-Ring, Part Number 79357
3. Lavage (Water) Filter, 10/Pack, Part Number 90158
For detailed information, contact your local Dentsply Sirona
Representative or authorized Dentsply Sirona Distributor.
9.6.3
Medicaments
The following medicaments are compatible with the Cavitron
Select SPS when used with the Cavitron Select Reservoir Pump:
• Chlorhexidine Gluconate 0.12%
(Oris™ CHX Antimicrobial Irrigating Solution)
• Cetylpyridinium Chloride (Cepacol®)
• Essential Oils (Listerine)
• Hydrogen Peroxide, 3% USP
• Povidone Iodine, 10% Solution
• Saline Solution
• Sanguinaria Extract (Viadent®)
• Sterile Water, USP
(For irrigating and System maintenance purposes).
10. System Setup, Operation and Techniques
for Use
10.1 Handpiece Setup
• This handpiece is sterilizable. Refer to Cavitron Systems
Infection Control Procedure Booklet for sterilization instructions
prior to using handpiece.
• Connect the Handpiece to the Cable Assembly by aligning the
electrical connections. If Cable Assembly does not seat into
the handpiece, gently rotate the handpiece until contacts align,
then fully insert handpiece.
• Hold empty handpiece in a semi-upright position over a sink
or drain. Press Foot Pedal until water exits to release any air
bubbles that might be trapped inside the handpiece.
• Lubricate the O-ring on the insert with water before placing it
into the handpiece. Fully seat insert with a gentle push-twist
motion. DO NOT FORCE.
• Turn the Lavage Control to select water flow rate during
system operation. Water flow rate is placed on a scale from 1 to
6 in older handpiece cable assemblies and small to large water
droplets in newer handpiece cable assemblies. Turn control
clockwise toward 6 or largest water droplet to increase water
flow at insert tip. Turn control counter-clockwise toward 1 or
smallest water droplet to decrease water flow. The water flow
rate through the handpiece also determines the temperature
of the lavage. Lower water flow rates produce warmer lavage.
Higher water flow rates produce cooler lavage. If the handpiece
becomes warm, increase the water flow rate. With experience
the Dental Healthcare Professional will be able to determine the
best water flow rate setting for optimum operating efficiency
and patient comfort.
10.2 Patient Positioning
For optimal access to both the upper and lower arches, the
backrest of the chair should be adjusted as for other dental
procedures. This assures patient comfort and clinician visibility.
Have the patient turn his/her head to the right or left. Also
position chin up or down depending upon the quadrant and
surface being treated. Evacuate irrigant using either a saliva
ejector or High Volume Evacuator (HVE).
10.3 Performing Ultrasonic Scaling Procedures
Note: Refer to Section 11 of this manual for general
procedures to be followed at the beginning of each day and
between patients.
• Follow precautions listed in the General and Ultrasonic sections
of 5.2 Procedural Precautions.
• The edges of Cavitron Ultrasonic Inserts are intentionally
rounded so there is minimal danger of tissue laceration with
proper ultrasonic scaling technique. Whenever the insert tip is
placed in the patient's mouth, the lips, cheek and tongue should
be retracted to prevent accidental (prolonged) contact with
the activated tip.
• Turn Power Level Control to select ultrasonic power level for
operation. Upward increases system power. Power level will
increase throughout the full range of the control. Hold the
handpiece over a sink or drain. Depress foot pedal to first
position to activate the system. Check water spray to verify
fluid is reaching the working end of the insert tip. Adjust the
water lavage control until the water (lavage) flows with a rapid
drip or small spray. Higher water flow settings provide cooler
irrigation.
• It may be necessary to adjust lavage with the system in "Boost"
mode (Foot Control fully depressed) so adequate fluid will be
available to cool tip to tooth interface.
• In general, it is suggested that a "feather-light-touch" be
used for ultrasonic scaling. The motion of the activated tip
and acoustic effects of the irrigating fluid, in most cases, are
adequate to remove even the most tenacious calculus.
• Periodically check the Cavitron Ultrasonic Insert for wear with
the Cavitron Insert Wear Indicator.
• The use of a saliva ejector or High Volume Evacuator (HVE) is
recommended during all procedures.
• Set the system's Power Level Control to the lowest efficient
power setting for the application and the selected insert.
10.4 Patient Comfort Considerations
Reasons for sensitivity
• Incorrect tip placement. The point should never be directed
toward tooth root surfaces.
• Not keeping tip in motion on tooth. Do not allow the insert to
remain in a static position on any one area of the tooth. Change
the insert's path of motion.
• Applying excessive pressure. Use a very light grasp and
pressure, with a soft tissue fulcrum whenever possible,
especially on exposed cementum.
If sensitivity persists, decrease power setting and/or move from
the sensitive tooth to another and then return.

11. System Care

It is recommended that you perform the following mainte-
nance procedures.
*NOTE: Refer to the Infection Control Procedure Booklet
supplied with your system for disinfectant products that
have been tested for compatibility
English - 12
Cavitron Select SPS Ultrasonic Scaler
Start-Up
1.
Open the manual shut-off valve on the dental
office water Supply system.
Procedures
2. Install a sterilized Steri-Mate Handpiece onto the
(beginning of
handpiece cable.
day)
3. Set the Power Level Control to minimum and
the Lavage Control to maximum.
4. Turn the system ON.
5. Hold the sterilized handpiece (without insert
installed) over a sink or drain. Activate the foot
control and flush the line for 2 minutes.
6. Place a sterilized 30K Cavitron Ultrasonic Insert
into the handpiece and set the Power knob
Control and Lavage Control to your preferred
operating position for ultrasonic scaling.
Between Patients
1.
Remove the used Cavitron Ultrasonic
Insert. Clean and sterilize Following the Infection
Control Procedures that were enclosed with
your insert
2.
Hold the handpiece over a sink or drain and
flush the waterline of the unit as above at maxi-
mum water flow for 30 seconds.
3.
Turn the System to the OFF position.
4.
Remove the Steri-Mate handpiece, clean and
sterilize following the procedures outlined in the
Cavitron Systems Infection Control Procedures
Booklet that was enclosed with your system.
5.
Disinfect the surfaces of the cabinet, Power
Cord, Handpiece Cable, Foot Control and Cable
assembly (if applicable), Water Supply lines as
instructed in the Cavitron Systems Infection
Control Procedures booklet. DO NOT SPRAY
DISINFECTANT SOLUTION DIRECTLY ON SYS-
TEM SURFACES*.
6.
Inspect the handpiece cable for any breaks or
tears.
7.
If using a closed water supply or DualSelect
Dispensing System, check for adequate fluid
volumes for the next patient.
8.
When ready to use, place a sterilized Steri-Mate
Handpiece onto the handpiece cable assembly
and insert a sterilized ultrasonic insert into the
handpiece and adjust system controls as pre-
ferred.
Shut-Down
Follow the "Between Patients" maintenance
procedures, Steps 1 through 6. If a quick disconnect
Procedures
is attached to the end of the hose:
(end of day)
1.
Disconnect the water supply line from the
dental office water supply.
2.
Relieve the water pressure by pressing the tip
of the connector into an appropriate container
and allow water to drain.
If quick-disconnect is not attached to the end of the
hose, it is recommended to close the manual shut
off valve on the dental water supply system.
Cavitron Select SPS Ultrasonic Scaler with
Cavitron Select Reservoir System
1.
Carefully lock Cavitron Select Reservoir Bottle into
Cavitron Select Reservoir Pump.
2.
Fill Cavitron Select Reservoir Bottle with tap, dis-
tilled water, or medicament. Replace Cavitron Select
Bottle cap.
3.
Connect Cavitron Select Scaler to Cavitron Select
Reservoir Pump and Bottle (see Section 8.5 for
complete instructions).
4.
Open the manual shut-off valve on the dental office
water supply system.
5.
Install a sterilized Steri-Mate Handpiece onto the
handpiece cable.
6.
Set the Power Level Control to minimum and the
Lavage Control to maximum.
7.
Turn the system ON.
8.
Hold the sterilized handpiece (without insert
installed) over a sink or drain. Depress the Foot
Control and flush the waterline for 2 minutes
9.
Place a sterilized 30K Cavitron Insert into the Hand-
piece. Set Power level knob and Lavage Control to
your preferred operating position for ultrasonic scal-
ing.
1.
Follow steps 1–6 listed in Between Patients (Cavitron
Select SPS Ultrasonic Scaler).
2.
With the power OFF, unscrew cap and fill Cavitron
Select Reservoir Bottle with tap, distilled water, or
medicament.
3.
Do not overfill Bottle with liquid into the neck of the
Bottle. Securely screw cap back on Bottle.
4.
When ready to use, place a sterilized Steri-Mate
Handpiece onto the handpiece cable assembly and
insert a sterilized ultrasonic insert into the hand-
piece and adjust system controls as preferred.
1.
Remove the used Cavitron Ultrasonic Insert. Clean
and sterilize following the Infection Control Proce-
dures that were enclosed with your insert.
2.
Turn the System OFF.
3.
Detach Cavitron Select SPS Scaler from Reservoir
System.
4.
Detach Cavitron Select Reservoir Bottle from Pump
(see Section 8 for complete instructions).
5.
Empty entire contents of Cavitron Select Reservoir
Bottle.
6.
Replace empty Cavitron Select Reservoir Bottle to
Pump.
7. Carefully lock Cavitron Select Reservoir Bottle into
Cavitron Select Reservoir Pump.
8.
Connect Cavitron Select SPS Scaler to Cavitron
Select Reservoir System (see Section 6 for complete
instructions).
9.
Turn the System ON.