Lubricate, Disinfect/Sterilize The Treatment Instruments; Treatment Instruments With Separate Instructions For Use; Cleaning And Sterilizing The Standard 3-Way Syringe - Dentsply Sirona Intego Operating Instructions Manual

Hide thumbs Also See for Intego:
Table of Contents

Advertisement

5 Care, cleaning and maintenance by the practice team
5.3 Instruments and instrument hoses
134°C 274°F
3 min
242
5.3.5

Lubricate, disinfect/sterilize the treatment instruments

5.3.5.1

Treatment instruments with separate instructions for use

The procedures required for the following treatment instruments are
described in the following separate instructions for use:
● Straight and contra-angle handpieces in various versions
● Turbines
● SiroSonic L ultrasonic handpiece
● Cavitron scaler
5.3.5.2

Cleaning and sterilizing the standard 3-way syringe

The Standard 3-way syringe can be cleaned under running water. Spray
and wipe disinfection can be used. The jet can be removed from the tip
H O
for sterilization.
2
➢ Press on the ring-shaped unlock button.
Ä The jet adapter is released. The jet will fall out.
To attach the jet, press it in until it clicks audibly into place.
The nozzle of the Standard 3-way syringe must lock into the holder.
Otherwise, the jet could fall out during treatment.
➢ Prior to treatment, ensure that the jet is securely in place.
WARNING
Dentsply Sirona
Operating Instructions Intego
64  5 7  4 98 D3543
D3543.201.01.08.02    02.2020

Advertisement

Table of Contents
loading

This manual is also suitable for:

Intego proIntego ambidextrousIntego pro ambidextrous

Table of Contents