5.3.4
Care for, disinfect and sterilize the treatment instruments
5.3.4.1
Treatment instruments
The procedures required for the following treatment instruments are
described in the following separate operating instructions:
● BL motor
● BL ISO E/C motor
● Straight and contra-angle handpieces in various versions
● Highspeed handpieces
● Sprayvit M multifunctional syringe
● SiroSonic TL scaler
5.3.4.2
Clean and disinfect/sterilize the components of the ApexLocator
NOTE
Approved care, cleaning, and disinfecting agents
Use only care, cleaning, and disinfecting agents that have been
approved by Dentsply Sirona, see "Care, cleaning, and disinfecting
agents" [→ 203].
Disconnect the mucosal electrode from the connection cable. The metal
hook can be sterilized, and the connection cable can be disinfected by
wiping.
134°C 274°F
3 min
The file clamp for manual measurement can be sterilized with the
connection cable.
134°C 274°F
3 min
The apex adapter and its connection cable can be disinfected by wiping.
IMPORTANT
To guarantee conductivity, no disinfectant must be allowed to
penetrate the electrical contacts.