Sterilization Instructions - Dentsply Sirona Tulsa 6:1 Operating Instructions Manual

Contra-angle handpiece
Table of Contents

Advertisement

274°F
134°C
62 39 920 D 3451
7.6

Sterilization Instructions

Prior to sterilization:
Clean and disinfect Contra-Angle per section 7.5.
P
Protect Contra-Angle by applying oil spray per section 7.3.
Professional
DAC
Contra-Angle handpieces may be sterilized only in the
steam sterilizer with saturated water vapor at 134 °C
(274 °F), 3 min.holding time (2.04 bar; 29.59 psi overpres-
sure).
Steam sterilizers may be used that fulfill the requirements of
EN 13060, class B or S and are suitable for the sterilization
The Hygiene Division of Sirona
The Hygiene Division of Sirona
of Contra-Angle handpieces.
ATTENTION! Do not exceed 140° C (284° F) even dur-
ing the drying phase!
7 Step-by-Step Instructions
20