Cleaning, disinfection and sterilization

IMPORTANT
Combine cleaning with disinfection
Manual cleaning must always be combined with disinfection.
All tissue residues must be removed from the optical fiber before it is
detached from the handpiece.
Following treatment, switch off the SiroLaser Blue and disconnect the
power cable from the power supply.
IMPORTANT
Gloves
Wear gloves during these procedures.
Control unit, handpiece body and handpiece tube must be cleaned and
wipe-disinfected.
Dispose the single-use fiber tips.
The removable stainless steal handpiece sleeve, therapy light guides,
fibercutter and bending tool must be cleaned and sterilized.
NOTICE
Parts using a washer-disinfector
Do not clean and disinfect the parts using a washer-disinfector! The
parts may be seriously damaged.
For the number of sterilization cycles, see chapter "Replacement of
parts subject to wear and tear  [→ 73]".
7 Cleaning, disinfection and sterilization
63